Starna References for Fluorescence Spectroscopy
Measurements by fluorescence spectroscopy are subject to many more instrumental and physical variables than absorption measurements. Due to these effects, the same sample may not give identical results with different instruments even if they are of similar design. Physical effects such as quenching or self-absorption may lead to departures from expected concentration/emission relationships over quite narrow concentration ranges. For these reasons it may be impossible to produce universal “absolute” reference materials, as certified values may be instrument dependent. Provided homogeneity and stability can be established, however, reference materials can be produced that allow users to monitor the performance of a spectrofluorometer over time and can also indicate the relative performance of different instruments. Starna fluorescence references are produced in an ISO17034 accredited production facility and calibrated in an ISO/IEC 17025 accredited calibration laboratory. Several of these references are cited in USP Chapter for spectrofluorometer qualification.
Spectral Response & Linearity
Starna Quinine Sulfate References
Used to qualify the spectral response of a spectrofluorometer and can also, over a limited concentration range (up to 1 mg/l), be used as a linearity check. USP cited reference.
Sensitivity & Signal to Noise Ratio
Starna High Purity Water Reference
This Reference Material uses Raman scattering to simulate the behaviour of fluorescent materials at very low concentrations and can be used to determine the Signal to Noise Ratio of a spectrofluorometer near its limit of detection and hence indicate its ultimate sensitivity. USP <853 > cited reference.
More information
Spectrofluorometer Performance and Wavelength Validation
Starna 6BF Series
Used for spectrofluorometer wavelength qualification, as a routine check of spectrofluorometer performance and for inter-instrument and inter-laboratory comparisons. Six polymer blocks containing seven organic fluorophores. Excitation range from 260 to 600 nm and emission range from 370 to 670 nm
More information
Phosphorescence Decay Rate and Fluorescence Lifetime
Starna SF Series Polymer Block References
Used to confirm measurement of transient fluorescentsignals. Polymer block doped with Europium III. Decay rate and phosphorescence lifetime certified for gate times of 50 μS and 2 μS and a measurement delay time of 0.1 mS using excitation and emission wavelengths of 340 nm and615 nm respectively.
More information
Microplate Fluorometer Qualification
Starna Fluorescence Reference Plates
For routine qualification of fluorescence-based plate readers for fluorescence intensity and well-to-well reproducibility. 96-well microplates containing inorganic fluorescent reference materials in glass matrix. The inorganic fluorophores are inherently stable, and do not suffer from photodegradation or bleaching.Formerly manufactured and supplied by Matech, USA. Available in top- and bottom-reading configurations. Range of excitation and emission wavelengths to match the user’s application.
More information
Starna Quantum Counter
Rhodamine B Solution Cell
Used by instrument manufacturers as a quantum counter to determine spectral response factors to calculate ‘corrected’ fluorescence spectra. Designed to provide unity quantum conversion, so that the emitted fluorescence is directly proportional to the intensityof the exciting radiation. Usable range 220 nm – 580 nm.
More information
Starna Quantum Yield Reference
Dilute Rhodamine 101 Solution Cell
Quantum yield of this solution is 1.0 at 25°C. Working excitation range 450 to 465 nm and emission range 550 to 650 nm. Blank-corrected absorbance of this solution is below thelevel recommended in the literature to avoid internal filter effects. Dilute Rhodamine 101 in ethanol in a sealed quartz cuvette with ethanol blank.
More information
Starna References for Fluorescence Spectroscopy
Rhodamine 101 Solution Cell
Starna References for Fluorescence Spectroscopy
Measurements by fluorescence spectroscopy are subject to many more instrumental and physical variables than absorption measurements. Due to these effects, the same sample may not give identical results with different instruments even if they are of similar design. Physical effects such as quenching or self-absorption may lead to departures from expected concentration/emission relationships over quite narrow concentration ranges. For these reasons it may be impossible to produce universal “absolute” reference materials, as certified values may be instrument dependent. Provided homogeneity and stability can be established, however, reference materials can be produced that allow users to monitor the performance of a spectrofluorometer over time and can also indicate the relative performance of different instruments. Such references are referred to as “Reference Materials” (RMs) or Quality Control Materials (QCMs). Starna produce a range of such materials under the same tight quality control procedures as are applied to their CRMs.
For the reasons outlined above, instrument qualification may only be valid for a particular instrument and within a relatively narrow range of operating parameters – indeed it can even be method specific. The references listed below will cater for many routine situations, but Starna can also design and supply fluorescence Reference Materials tailored to individual needs.
Quinine Sulfate Solution References
used to calibrate the spectral response of a spectrofluorometer and as a linearity check.
High Purity Water (Raman) Reference
used to determine the signal-to-noise ratio of a fluorescence spectrophotometer and hence its ultimate sensitivity.
Doped Polymer Blocks (Starna SF Series)
spectral response references, traceable to NIST Fluorescence SRMs NEW!
Europium III Phosphorescence Reference
used to confirm measurements of transient fluorescentsignals.
Starna Rhodamine B Quantum Counter
used in fluorescence instrumentation for spectral correction
Starna Quantum Yield Reference
used to determine the fluorescence quantum yield of a sample
Doped Polymer Blocks (Starna 6BF Series)
for spectrofluorometer wavelength qualification andas a routine check of spectrofluorometer performance.
Rare Earth Solution Cells
for spectrofluorometer wavelength qualification
Doped Glass Microplate References
qualification of fluorescence-based plate readers for fluorescence intensity and well-to-well reproducibility NEW!