Measurements by fluorescence spectroscopy are subject to many more instrumental and physical variables than absorption measurements. Due to these effects, the same sample may not give identical results with different instruments even if they are of similar design. Physical effects such as quenching or self-absorption may lead to departures from expected concentration/emission relationships over quite narrow concentration ranges. For these reasons it may be impossible to produce universal “absolute” reference materials, as certified values may be instrument dependent. Provided homogeneity and stability can be established, however, reference materials can be produced that allow users to monitor the performance of a spectrofluorometer over time and can also indicate the relative performance of different instruments. Starna fluorescence references are produced in an ISO17034 accredited production facility and calibrated in an ISO/IEC 17025 accredited calibration laboratory. Several of these references are cited in USP Chapter for spectrofluorometer qualification.


Spectral Response & Linearity
Starna Quinine Sulfate References 

Used to qualify the spectral response of a spectrofluorometer and can also, over a limited concentration range (up to 1 mg/l), be used as a linearity check. USP cited reference.

More information 


Sensitivity & Signal to Noise Ratio
Starna High Purity Water Reference

This Reference Material uses Raman scattering to simulate the behaviour of fluorescent materials at very low concentrations and can be used to determine the Signal to Noise Ratio of a spectrofluorometer near its limit of detection and hence indicate its ultimate sensitivity. USP <853 > cited reference.

More information


Spectrofluorometer Performance and Wavelength Validation
Starna 6BF Series

Used for spectrofluorometer wavelength qualification,  as a routine check of spectrofluorometer  performance and for inter-instrument and inter-laboratory comparisons. Six polymer blocks  containing seven organic fluorophores. Excitation range from 260 to 600 nm and emission range from 370 to 670 nm

More information


Phosphorescence Decay Rate and Fluorescence Lifetime
Starna SF Series Polymer Block References

Used to confirm measurement of transient fluorescentsignals. Polymer block doped with Europium III. Decay rate and phosphorescence lifetime certified for gate times of  50 μS and 2 μS and a measurement delay time of 0.1 mS using excitation and emission wavelengths of 340 nm and615 nm respectively.

More information


Microplate Fluorometer Qualification
Starna Fluorescence Reference Plates

For routine qualification of fluorescence-based plate readers for fluorescence intensity and well-to-well reproducibility.  96-well microplates containing inorganic fluorescent reference materials in glass matrix. The inorganic fluorophores are inherently stable, and do not suffer from photodegradation or bleaching.Formerly manufactured and supplied by Matech, USA. Available in top- and bottom-reading configurations. Range of excitation and emission wavelengths to match the user’s application.

More information


Starna Quantum Counter
Rhodamine  B Solution Cell

Used by instrument manufacturers as a quantum counter  to determine spectral response factors to calculate ‘corrected’ fluorescence spectra. Designed to provide unity quantum conversion, so that the emitted fluorescence is directly proportional to the intensityof the exciting radiation. Usable range 220 nm – 580 nm. 

More information


Starna Quantum Yield Reference
Dilute Rhodamine 101 Solution Cell

Quantum yield of this solution is 1.0 at 25°C. Working excitation range 450 to 465 nm and emission range 550 to 650 nm. Blank-corrected absorbance of this solution is below thelevel recommended in the literature to avoid internal filter effects. Dilute Rhodamine 101 in ethanol in a sealed quartz cuvette with ethanol blank.

More information


 

Measurements by fluorescence spectroscopy are subject to many more instrumental and physical variables than absorption measurements. Due to these effects, the same sample may not give identical results with different instruments even if they are of similar design. Physical effects such as quenching or self-absorption may lead to departures from expected concentration/emission relationships over quite narrow concentration ranges. For these reasons it may be impossible to produce universal “absolute” reference materials, as certified values may be instrument dependent. Provided homogeneity and stability can be established, however, reference materials can be produced that allow users to monitor the performance of a spectrofluorometer and make inter-instrument and inter-laboratory comparisons. Starna fluorescence references are produced in an ISO17034 accredited production facility and calibrated in an ISO/IEC 17025 accredited calibration laboratory. Several of these references are cited in USP Chapter <853> for spectrofluorometer qualification. 

 

Quinine Sulfate References -Spectral Response and Linearity 

One of the first reference materials developed for fluorescence spectroscopy. Equivalent to NIST SRM 936a and cited in USP <853 >.

 

High Purity Water Reference- Sensitivity & Signal to Noise Ratio 

Used to determine the Signal to Noise Ratio of a spectrofluorometer near its limit of detection and hence indicate its ultimate sensitivity. USP <853 > cited reference.

 

Starna SF Series Doped Polymer References - Spectral Response Qualification

Five polymer blocks containing stable organic fluorophores.  Excitation range 280 to 700 nm and emission range 380 to 770 nm

 

Starna 6BF Series Doped Polymer References -Wavelength Validation and Performance Check

Six polymer blocks containing seven organic fluorophores. Excitation range 260 to 600 nm and emission range 370 to 670 nm.

 

Starna Europium III Reference - Phosphorescence Decay Rate and Fluorescence Lifetime  

Polymer block doped with Europium III. Decay rate and phosphorescence lifetime certified for gate times of 50 μS and 2 μS.  

 

Rhodamine Solution Cell - Quantum Counter

Used by instrument manufacturers to determine spectral response factors to calculate ‘corrected’ fluorescence spectra. Designed to provide unity quantum conversion,

 

Rhodamine in Ethanol Cell - Quantum Yield Reference

Quantum yield 1.0 at 25°C. Blank-corrected absorbance of this solution is below the level recommended in the literature to avoid internal filter effects. 

 

Starna Rare Earth Solution Cells -Spectrofluorometer Wavelength Qualification 

Cited in USP Chapter <853> for spectrofluorometer wavelength qualification. 

 

Doped Glass Microplate References - Microplate Fluorometer Qualification 

For qualification of fluorescence-based plate readers. 96-well microplates containing inorganic fluorescent reference materials in a glass matrix. Formerly supplied by Matech, USA.

 

Measurements by fluorescence spectroscopy are subject to many more instrumental and physical variables than absorption measurements. Due to these effects, the same sample may not give identical results with different instruments even if they are of similar design. Physical effects such as quenching or self-absorption may lead to departures from expected concentration/emission relationships over quite narrow concentration ranges. For these reasons it may be impossible to produce universal “absolute” reference materials, as certified values may be instrument dependent. Provided homogeneity and stability can be established, however, reference materials can be produced that allow users to monitor the performance of a spectrofluorometer over time and can also indicate the relative performance of different instruments. Such references are referred to as “Reference Materials” (RMs) or Quality Control Materials (QCMs). Starna produce a range of such materials under the same tight quality control procedures as are applied to their CRMs. 

For the reasons outlined above, instrument qualification may only be valid for a particular instrument and within a relatively narrow range of operating parameters – indeed it can even be method specific. The references listed below will cater for many routine situations, but Starna can also design and supply fluorescence Reference Materials tailored to individual needs.  


Quinine Sulfate Solution References

used to calibrate the spectral response of a spectrofluorometer and as a linearity check. 


High Purity Water (Raman) Reference

used to determine the signal-to-noise ratio of a fluorescence spectrophotometer and hence its ultimate sensitivity. 


Doped Polymer Blocks (Starna SF Series)

spectral response references, traceable to NIST Fluorescence SRMs NEW! 


Europium III Phosphorescence Reference

used to confirm measurements of transient fluorescentsignals.  


Starna Rhodamine B Quantum Counter

used in fluorescence instrumentation for spectral correction 


Starna Quantum Yield Reference

used to determine the fluorescence quantum yield of a sample 


Doped Polymer Blocks (Starna 6BF Series)

for spectrofluorometer wavelength qualification andas a routine check of spectrofluorometer performance. 


Rare Earth Solution Cells

for spectrofluorometer wavelength qualification 


Doped Glass Microplate References

qualification of fluorescence-based plate readers for fluorescence intensity and well-to-well reproducibility NEW!