Tools for Easier Compliance

The regulations for ultraviolet and visible spectroscopy were recently updated by the US and European pharmacopoeias. The revised US Pharmacopeia (USP) Chapter became mandatory on 1st December 2019. European Pharmacopoeia 10.0 (EP) Chapter 2.2.25 became mandatory on 1st January 2020.

Both standards now require absorbance linearity to be qualified and that qualification measurements are made at parameter values that match or ‘bracket’ those used for analysis. This may well mean that a single reference for e.g. stray light will no longer suffice and explains why potassium chloride solution, for example, is no longer a universal stray light reference but is recommended for the far UV. It also means that it is no longer possible to specify universal ‘kits’ that can qualify instruments for all applications.

The Basic Qualification Kits described below, however, will meet the routine needs of many laboratories and can be supplemented with additional references if required. Starna CRM kits offer cost savings when compared to the purchase of individual references.

Basic Qualification Kits

The kits below will allow instrument qualification over the stated wavelength and absorbance ranges for compliance with EP 2.2.25 and USP . References are provided to qualify wavelength accuracy, absorbance accuracy and 3-point linearity, stray light and resolution/bandwidth.

The stray light references for USP compliance are for Chapter Method B. Note that Potassium Chloride solution is included in all the basic kits as even if the instrument is not being used in the far UV, increasing stray light at this wavelength (200nm) is a sensitive indication of deterioration of the instrument optics.

These kits have been designed to qualify UV-visible spectrophotometers over the wavelength and absorbance ranges commonly used in pharmaceutical analysis. If your operating parameters fall outside these ranges, Starna has the world’s biggest range of spectroscopy CRMs and will be pleased to recommend additional or alternative references to meet individual requirements. For more information see Selecting References for Compliance - a User Guide or contact sales@starna.com for advice.

Basic Kit for EP 2.2.25 compliance 
Catalogue number: RM-EP20

[Kit identity: RM-020610HLKCSIKISNTX]

Reference

To qualify

Wavelength range

Holmium cell

Wavelength accuracy

240 - 650 nm

Potassium dichromate 20, 60 & 100 mg/l & blank cell.

Absorbance accuracy and linearity, 0-1 A

 

235 – 350 nm

Potassium chloride

Stray light, 200nm

200 -370 nm

Sodium iodide 

Stray light, 220 nm

Potassium iodide

Stray light, 250 nm

Sodium nitrite

Stray light, 340, 370 nm

Toluene in hexane

Resolution (bandwidth), 265-270nm

265-270 nm

 

Basic kit for USP Chapter <857 compliance  

Catalogue number RM-USP20

[Kit identity RM-02061060HLKCKIACSNTX]                                                              

Reference

To qualify

Wavelength range

Holmium cell

Wavelength accuracy

 

Potassium Dichromate
20, 60 & 100 mg/l & blank cell.

Absorbance accuracy and linearity, 0-1 A

235 – 350 nm

Potassium dichromate

Absorbance accuracy, 1 A

430 nm

Potassium chloride

Stray Light, 200 nm

240 - 370 nm

Potassium iodide

Stray light, 220 nm

Acetone

Stray light, 300 nm

Sodium nitrite

Stray light, 340 nm

Toluene in hexane

Resolution (bandwidth), 265-270nm

265-270 nm

 

Basic Combination kit for USP and EP compliance
Catalogue number RM-USPEP20
[Kit identity RM-02061060HLKCSIKIACSNTX]

 Reference

To qualify

Wavelength range

Holmium cell

Wavelength accuracy

240 - 650 nm

Potassium Dichromate 20, 60 & 100 mg/l & blank cell.

Absorbance accuracy and linearity, 0-1 A

 

235 – 350 nm

Potassium dichromate

Absorbance accuracy, 1 A (USP)

430 nm

Potassium chloride

Stray light, 200 nm (EP &USP)

200 -370 nm

Sodium Iodide 

Stray light, 220 nm (EP)

Potassium Iodide

Stray light, 220 nm (USP), 250nm (EP)

Acetone

Stray light, 300 nm (USP)

Sodium Nitrite

Stray light, 340 nm (EP & USP), 370 nm (EP)

Toluene in Hexane

Resolution (bandwidth), 265-270nm

265-270 nm

 

 

Far UV extension kit - EP and USP compliance
Catalogue number: RM-06A12A24ARE

Reference

To qualify

Wavelength range

Cerium cell 

Wavelength accuracy

200 -270 nm

Nicotinic acid 6, 12 & 24 mg/l

Absorbance accuracy and linearity, 0-1 A 

210 - 270 nm

 

 

Visible Absorbance and Wavelength kit 
Catalogue number RM-1N3N5DHG

Reference

To qualify

Wavelength range

Holmium Filter 

Wavelength accuracy

240 – 640 nm

Neutral Density Filters, 0.25, 0.50 & 1.00A

Absorbance Accuracy and Linearity

440 – 635 nm